Product details
Product Overview
- FC-R7700 is an analog mass flow controller from AERA.
- It belongs to the FC-R7700 Series with elastomer seals.
- The unit covers 10 sccm to 5 slm full scale.
- Accuracy stays within +/-1% of full scale.
- Repeatability is better than +/-0.2% of full scale.
- Response time is under 2 seconds to +/-2% of full scale.
- This model suits non-corrosive inert gases such as N2, O2 and Ar.
- The design is field-proven and cost-effective for industrial users.
- It replaces many other analog MFC brands with little effort.
- Typical uses include CVD, PVD, etch, diffusion and ion implantation.
- It also fits flat-panel display, data storage and coating lines.
Technical Data
- Brand: AERA
- Model: FC-R7700
- Series: FC-R7700 Series (analog, elastomer-sealed)
- Module type: Analog mass flow controller
- Full-scale flow range: 10 sccm to 5 slm
- Accuracy: +/-1% of full scale or better
- Repeatability: +/-0.2% of full scale or better
- Linearity: +/-0.5% of full scale or better
- Response time: below 2 seconds settling time
- Control range: 2% to 100% of full scale
- Zero stability: under 0.5% of full scale per year
- Gas type: all inert, non-corrosive gases
- Operating pressure: differential 7 to 40 psiD
- Normal operating pressure: 70 psig
- Maximum inlet pressure: 140 psig
- Wetted materials: 316(L) stainless steel, PTFE, magnetic stainless
- Seals: fluoroelastomer or chloroprene rubber
- Valve type: solenoid, normally-open or normally-closed
- Leak integrity: 1 x 10^-6 atm-cc/sec He max
- Electrical interface: analog 0 to 5 VDC command and output
- Connector: Aera 9-pin D connector
- Power supply: +15 VDC at 25 mA and -15 VDC at 180 mA
- Power draw: 3.1 W max
- Operating temperature: 5 to 45 degrees C (41 to 113 degrees F)
- Heated option: none, standard unheated version
- Dimensions: 119 x 32 x 124 mm (H x W x D)
- Weight: 1.0 kg (2.2 lb)
Measurement and Signal Performance
- Flow range: 10 sccm to 5 slm full scale
- Frequency response: flow settles in under 2 seconds
- Sensitivity: low full-scale error supports fine control
- Linearity: within +/-0.5% of full scale
- Resolution: stable control from 2% to 100% of range
- Temperature range: 5 to 45 degrees C
- Environmental tolerance: works in clean, non-corrosive gas lines
- Noise immunity: analog 0 to 5 VDC link with D connector
Installation Guide
- Mount the body on a stable, level surface.
- Keep the 32 mm body width clear for airflow.
- Use VCR, VCO or Swagelok-compatible fittings.
- Check differential pressure stays between 7 and 40 psiD.
- Keep inlet pressure below 140 psig at all times.
- Connect the 9-pin D plug to the analog controller.
- Feed clean +15 VDC and -15 VDC power.
- Purge the line with dry inert gas before use.
Setup and Maintenance
- Zero the unit after warm-up for best stability.
- Check zero drift once a year.
- Replace elastomer seals based on gas exposure.
- Keep the wetted path free of corrosive gas.
- The analog interface makes bench checks simple.
- Quick swap design cuts downtime on replacement.
Compatibility Notes
- Fits VCR, VCO and Swagelok-style connections.
- Drops into many existing analog MFC slots.
- Works with standard 0 to 5 VDC control loops.
- Replaces other brands with minor setup changes.
- Use only with inert, non-corrosive gases.
Certification and Compliance
- RoHS compliant for restricted substances.
- Leak integrity meets 1 x 10^-6 atm-cc/sec He max.
- Leak rate also meets 1 x 10^-7 Pa-m3/sec He max.
Packing List
- One AERA FC-R7700 mass flow controller
- One 9-pin D connector kit
- One calibration certificate
- One quick start guide
FAQ
- Q: Which gases can I run through this unit?
A: Use inert, non-corrosive gases only. Examples are N2, O2 and Ar. - Q: How fast does the flow settle?
A: It settles within 2 seconds to +/-2% of full scale. - Q: Can I use it in a CVD tool?
A: Yes. CVD, PVD, etch and diffusion are common uses. - Q: Does it need a heated body?
A: No. This is the standard unheated version. - Q: What power does it need?
A: It needs +15 VDC and -15 VDC, with 3.1 W max draw. - Q: How often should I check zero?
A: Check once a year. Drift stays under 0.5% of full scale. - Q: Will it fit my current gas line?
A: It accepts VCR, VCO and Swagelok-compatible fittings. - Q: From experience, what causes poor accuracy?
A: Dirty gas and wrong differential pressure are common causes. Clean the line and recheck 7 to 40 psiD. - Q: What is a practical tip for seal life?
A: Keep corrosive gas out of the line. Swap elastomer seals on a fixed schedule. - Q: Can I replace another brand without a full rebuild?
A: Often yes. The 0 to 5 VDC link and standard fittings ease the swap.

















There are no reviews yet.